Channel thresholds:
|
0.1, 0.2, 0.3, 0.5, 1.0, and 5.0 μm
|
Flow rate:
|
1 CFM (28.3 LPM); monitored and adjusted electronically
|
Calibration:
|
Materials traceable to US National Institute for Standards and Technology (NIST)
|
Test methodology:
|
0.1 μm particles verified by Differential Mobility Analysis; larger sizes verified by transmission electron microscopy
|
Maximum concentration:
|
@ 5% coincidence loss: > 500,000/ft3 (optional diluter required for ISO Class 8+); automatic over-coincidence alarm
|
Zero count level:
|
Exceeds JIS standards, allowing use in ISO class 1 and 2 rooms
|
Laser source:
|
Laser diode, 3W (de-rated to 2W), 810 nm, infrared
|
Laser cooling:
|
Thermoelectric chilling (Peltier cooler), heat sink, dedicated 5 CFM cooling fan
|
Warranty:
|
Three year warranty on laser module, two years on other components
|
Communication modes:
|
Ethernet or RS-232; optional 4-20 mA output board; optional USB output
|
Environmental sensors:
|
Optional Temp/RH probe; integrates up to 4 external sensors via 4-20 mA input
|
Display:
|
1/4 VGA color LCD screen
|
Display languages:
|
English, French, German, Italian, Japanese (Kanji), Korean, Spanish, Mandarin (classical or simplified)
|
Printer:
|
Built-in thermal printer
|
Reports:
|
Sample print strips; final reports for ISO/FS209/British Standards cleanroom qualification; averages
of multiple runs; print-on-alarm; sampling configuration printouts
|
Remote operation:
|
Use web browser to operate unit; save or load recipes
|
Key software features:
|
Recipes, scheduled sampling, historical data filtering, sorting and review, password protection
|
Data storage:
|
3,000 data sets (includes particle and environmental data, plus location, time, etc.)
|
Data and sampling security:
|
Stored data records cannot be altered; sampling recipes can be password protected
|
External surface:
|
KYDEX chassis, polyester display screen, and zinc-plated steel bottom are all ESA/ESD compliant
|
Cleaning materials:
|
Bleach, formaldehyde, ethyl/isopropyl alcohol, peroxide/quaternary ammonium solutions
|
Monitor cleanliness:
|
Internally filters both pump exhaust (99.99+% @ 0.1 μm) and cabinet exhaust (95.0% @ 0.3 μm)
|
Dimensions (l, w, h);
weight:
|
18 x 14 x 13 in. ( 45 x 35 x 32 cm);
37 lb. (17 kg)
|
Power:
|
85-264 V, 50-60 Hz; optional external battery/charger
|
Operating environment:
|
Temperature: 15 - 30°C; Humidity: 10 - 85% non-condensing
|