Model
|
19071
|
19073
|
功能
|
ACV
|
ACV/DCV/IR
|
Withstanding
Voltage Test
耐压测试
|
Output Voltage
|
AC : 0.05 ~ 5kV
|
AC : 0.05˜ 5kV,
DC : 0.05 ~ 6kV
|
Load Regulation
|
1% + 5V
|
Voltage Resolution
|
2 V
|
Voltage Accuracy
|
1% + 5counts
|
Cutoff Current
|
AC : 0.1mA~ 20mA
|
AC : 0.1mA ~ 20mA,
DC : 0.01mA ~ 5mA
|
Current Resolution
|
AC : 1μA, DC : 0.1μA
|
Current Accuracy
|
1.5% + 5counts
|
Current Frequency
|
50Hz/ 60Hz
|
Test Time
|
0.1~999 sec, continue
|
Ramp up Time
|
0.1~999 sec, off
|
Waveform
|
Sine wave
|
Insulation
Resistance
Test
绝缘测试
|
Output Voltage
|
-
|
DC : 50 ~ 1000 V
|
Voltage Resolution
|
-
|
2V
|
Voltage Accuracy
|
-
|
5% + 5counts
|
IR Range
|
-
|
0.1MΩ ~ 50GΩ
≧ 500V : 1MΩ ~ 1000MΩ
± (4 % + 5 counts),
1GΩ ~ 10GΩ± (7 % + 5 counts),
|
Resistance Accuracy
|
-
|
10GΩ ~50GΩ± (12 % + 5 counts),
|
< 500V : 0.1MΩ ~1000MΩ
± (7 % + 5 counts)
|
< 100V : 0.1MΩ ~1000MΩ
± (10 % + 5 counts)
|
ARC
Detection
|
Setting Mode
|
Programmable setting
|
Detection Current
|
AC : 1mA ~ 15mA, DC : 1mA ~ 5mA
|
Minimum Pulse Width
|
10 μs approx.
|
Secure
Protection
Function
|
Fast Output Cut-off
|
Approx. 0.4mS, after NG happen
|
Fast Discharge
|
Approx. 0.2S, Typical
|
Ground Fault Interrupt
|
0.5mA ± 0.25mAac (ON), OFF
|
Continuity Check
|
0.1Ω ~ 5.0Ω ± 0.2Ω, GC MODE
|
Panel Operation Lock
|
Yes
|
GO/NG
Judgment
Window
|
Indication, Alarm
|
GO: Short sound; NG: Long sound
|
Data Hold
|
Least tests data memories
|
Step Hold
|
Step signal trigger ON / OFF
|
Memory Storage
|
10 tests setups with 60 steps pre setup
|
General
|
Operation Environment
|
Temperature: 0˚C ~ 40 ˚C, Humidity: ≦ 80 % RH
|
Power Consumption
|
No load : <60 W, With rated load : ≦ 300 W
|
Power Requirement
|
100V / 120V / 220V / 240V, 50 / 60 Hz
|
Dimension (H x W x D)
|
105x272.8x350mm/
4.13x10.74x13.78inch
|
105x270x350mm/
4.13x10.63x13.78inch
|
Weight
|
11 kg / 24.23 lbs
|
Certification
|
UL, TUV, CE
|