Model
|
VOC-1000
|
VOC-2000
|
Compact chamber
|
Temperature control range
|
+20 to +100°C :
At test operation (no load, with specimen)
|
+80 to +250°C :
Heat processing operation time (no load, no specimen)
|
Temperature and humidity control range
|
30 to 90%RH (at+20 to +50°C)(no load, with specimen)
|
Temperature and humidity control width
|
±0.5°C / ±5% (no load, no specimen)
|
Temperature and humidity distribution width*1
|
±1.0°C / ±5% (no load, no specimen)
|
Background concentration
|
Less than 2 µg/m³ for single substance concentration
|
Less than 20 µg/m³ for TVOC concentration
|
Air exchange rate
|
0.2 to 2.0 times per hour (at test operation)
|
0.2 to 1.5 times per hour (at test operation)
|
Air exchange rate control width
|
±5% (air exchange rate of 0.5 times per hour or higher)
|
Mass transfer coefficient
|
9 to 18m/h (water vapor conversion)
|
Air tightness
|
Air leakage under 1% of supply air amount (at air exchange rate of 0.5 times per hour)
|
Recovery rate
|
80% or higher (Toluene)
|
Effective inner dimension
(W × H × D mm)
|
1000 × 1100 × 810
|
2000 × 1100 × 810
|
(Inner capacity: Approx. 1 m³)
|
(Inner capacity: Approx. 2 m³)
|
Material (Configuration)
|
Stainless steel plate (full circumference welded structure)
|
Door effective dimension (W × H mm)
|
1000 × 1100
|
2000 × 1100
|
Door lock
|
Latch lock (6 locations)
|
Latch lock (16 locations)
|