AC/DC Hipot
|
AC Output Voltage:
|
Range: 50V to 5000V AC, 1V resolution
|
Regulation: <(1% of setting +5V) at Rated Load
|
Freq: 50-600 Hz, programmable in 1Hz steps
|
50~60Hz : 1% + 5 cnts, > 60Hz: 3%+10 cnts
|
Waveform: Sinusoidal
|
Voltage Display:
|
Accuracy: ±(1% of reading + 5V)
|
Resolution: 1Volt
|
AC Current Display:
|
Total current, Range: 0.001 to 40mA AC
|
Resolution: 1 or 10μA steps
|
Accuracy: ±(1% of reading + 5cnt)
|
High/Low Limit Test:
|
1μA to 40mA AC
|
Low limit can be turned OFF
|
Arc Detection:
|
Programmable 1-20mA,
|
Pulse Width 40μ, 20μ, 10µ or 4μsec
|
DC Output Voltage:
|
Range: 50V to 6000V DC, 1V resolution
|
Regulation: <(1% of setting +5V) at Rated Load
|
Voltage Display:
|
Accuracy: ±(1% of reading + 5V)
|
Resolution: 1Volt
|
DC Current Display:
|
Range: 0.1µ to 12mA DC
|
Resolution: 0.1, 1 or 10μA steps
|
Accuracy: ±(1% or reading + 5cnt)
|
High/Low Limit Test:
|
0.0001mA to 12mA DC
|
Low limit can be turned OFF
|
Arc Detection:
|
Programmable 1-10mA
|
Pulse Width 40μ, 20μ, 10µ or 4μsec
|
Insulation Resistance
|
Voltage Range:
|
50V to 1000V DC
|
Voltage Accuracy:
|
±(1% of setting + 5V)
|
High/Low Limit Test:
|
100kV - 50GV
|
Low limit can be turned OFF
|
Ground Bond:
|
Output Current:
|
Range: 1.0 to 30.0A AC, setting 0.1A/step; *40A
|
Accuracy: ±(1% of setting + 0.3A)
|
Frequency:
|
50 or 60Hz Selectable
|
No Load Voltage:
|
6 to 15 V Programmable
|
Resistance:
|
Range: 0.1mΩ-510.0mV, 4 digits; Resol: 1mV
|
Accuracy: ±(1% of reading + 3 counts)
|
High Limit:
|
0.1mΩ to 510mΩ
|
Start Wait:
|
0.1 - 99.9sec, OFF
|
Leakage Current:
|
Input Voltage:
|
Range: 0V to 300V
|
Line Voltage Meter:
|
0 -300V AC, ±(1% of reading + 6cnts)
|
Line Current Meter:
|
0 - 20A, ±(1.5% of reading + 0.1A)
|